E-mail

Acrel AID150 Alarm And Display Device

AID150 centralized alarm and display instrument adopts LCD liquid crystal display and RS485 bus, which can centrally monitor the data of up to 16 sets of AIM-M10 medical intelligent insulation monitors, and can remotely sound and light alarm. AID150 can also monitor the data of multiple sets of AIM-R100 residual current monitors. It is installed embedded in the wall and is suitable for centralized monitoring in operating rooms, intensive care units or other medical places.

isolation panel for operating rooms
isolation panel for operating rooms
operating room electrical panels
digital power meter price
electric digital meter price
Products

Specification of AID150 Alarm And Display Device

Parameters
Parameters/Model
AID10AID120AID150
Auxiliary powerVoltageDC 24V
Power consumption< 0.6W
Insulation resistance measurement range0—999kΩ
Insulation alarm value range
50—999kΩ
Transformer load rate displayPercentage display
Load current alarm settings14A, 18A22A28A35A45A
Temperature alarm value range0~+200℃
AlarmAudible and visible alarm
Alarm TypeInsulation fault, overload, over-temperature, equipment failure
CommunicationRS485 interface, MODBUS-RTU protocol
Display methodLED Display128*32 LCD


Features of AID150 Alarm And Display Device

01

Through the bus communication technology, the monitoring data and alarm information of the insulation monitor can be read and displayed in real time;

02

The alarm parameters of the insulation monitor can be set remotely;

03

When the system fails, it can sound and light alarm. After pressing the mute button, the sound will be turned off, and the corresponding fault indicator light will not go out until the fault is resolved.

Diagram of AID150 Alarm And Display Device

Diagram of AID150 Alarm And Display Device
Features
Diagram of AID150 Alarm And Display Device
Features
Diagram of AID150 Alarm And Display Device
Features
Diagram of AID150 Alarm And Display Device
Display Sample
Diagram of AID150 Alarm And Display Device
Wiring Instruction
Diagram of AID150 Alarm And Display Device
Typical Connection
Innovation, Efficiency, Unity, Integrity.
Acrel, your reliable energy meter partner. Don't hesitate to contact us!
Contact Us
Related Medical IT Insulation Monitoring System
+8618795636361 michelle.zhou@email.acrel.cn
No. 9 Xincheng Road, Chengjiang Street, Jiangyin City, Jiangsu Province, China.